AZO® UTI Test Strips are used as an aid in the screening of a Urinary Tract Infection (UTI). The UTI test strips are for the detection of Leukocytes (white blood cells) and Nitrite in urine. See insert inside packaging for complete instructions.
Directions Summary:
- Simply wet one test trip by holding it in your urine stream for 1-2 seconds.
- Read result at 1 minute for Nitrite and at 2 minutes for Leukocytes.
- Match the color of the test strip pads to the color blocks on the foil pouch.
LIMITATIONS OF USE: There is the possibility that this test may produce false results. Consult your physician before making any medical decisions. Note: The test may be affected by substances that cause abnormal urine color such as riboflavin (Vitamin B2) or products containing dyes as the active ingredient; for example phenazopyridine hydrochloride, the active ingredient in AZO® Urinary Pain Relief. The color development on the test pad may be masked or a color reaction may be produced that could be interpreted as false results. Reading of the results may be affected if you have color vision deficiency (color blindness).
Leukocytes: The drug tetracycline may cause a false negative result. High protein or elevated glucose in urine may cause test results to be low. Urine with a pH≥8.0 may cause a false negative leukocyte reading.
Nitrite: Any degree of uniform pink color should be read as a positive result. Pink spots or pink edges should not be read as a positive result. High ascorbic acid may cause test results to be low. Note: You may get a negative result if you have a UTI caused by bacteria that does not change nitrate to nitrite, when urine has not been held in the bladder for more than 4 hours, when taking antibiotics, or when your diet does not include nitrates.
WARNINGS AND PRECAUTIONS: Please read all the information in this package insert before performing the test.
- For urine test only. Do not use for blood testing.
- Do not use after the expiration date.
- Keep out of the reach of children.
- For in-vitro diagnostic use. Not to be taken internally.
- The used test strip, open foil pouch and drying packet should be discarded according to local regulations after testing.
STORAGE AND HANDLING:
- Store in a dry place at room temperature at 59-86° F or 15-30° C. Do not refrigerate. Keep out of direct sunlight.
- The test strip should remain in the foil pouch with the drying packet until use.
- Do not touch the test pads on the strip. Discard any discolored strips that may have deteriorated.